LDC-5/R
The REFERENCE CALIBRATOR LDC-5/R generates well producible charge pulses of high accuracy where the pulse magnitude can be adjusted continuously between 1 pC and 2000 pC.
Applications
- Generation
of reproducible charge pulses of high accuracy
- Pulse
magnitude can be adjusted continuously between 1 pC and 2000 pC
- Calibrating
of PD calibrators
- A complex
calibrating system, consisting of Pulse Integrator LDC-5/E and a precision
pulse peak detector can also be ordered
User Benefits
- Contains a
pulse amplifier and a peak detector
- Possibility
to connect a scope to monitoring PD events (DC coupling and a sensitivity of
0.5 V/div is reccomended)
- Powered by
: 9 V battery (6LR61), current consumption less than 10 mA
- BNC-Adapter,
fibre optic cable 10m
Specifications
- charge continuously adjustable from 1 pC to 2,000 pC
- resolution of the pre-selected charge values < 3 % of the selected measuring range
- calibration capacitor 100 pF
- battery-powered, battery "LOW" indicator
-